Support

FAQ

November 28, 2022

OBA measurement - instrument comparison experiment

In this experiment, the T42* substrate that contains OBAs is measured with a number of different Instruments. The following table compares the differences between these measurement devices related to measuring samples with OBAs:

 Instrument

OBA

Index

Fluorescence

Index

M1-M2 Spectral ∆ 

 @ 430 nm

M1-M2

 ∆E00 

i1 Pro 3  8.8 9.5 0.24@ 430 nm 8.71
i1 Pro 2  7.2 7.4 0.21@ 430 nm 7.24
eXact (v.1) M1 Part 2 export on 5.8 5.8 0.16 @ 430 nm 5.72
eXact(v.1) M1 Part 2 export off 6.2 5.9 0.15 @ 430 nm 6.08
Techkon SpectroDens 7.9 7.4 0.16 @ 440 nmn 7.25
NIX Spectro 2  soon      

 

A fast visual comparison of spectral curves explains why numbers are very different.

 Instrument Reflective Spectrum (M1, M0, M2)
i1 Pro 3 
i1 Pro 2 
eXact M1 Part 2 export on
eXact M1 Part 2 export off 
Techkon SpectroDens
NIX Spectro 2  soon

 

 Instrument Fluorescent Effect (M1-M2)
i1 Pro 3 
i1 Pro 2   
eXact M1 Part 2 export on
eXact M1 Part 2 export off 
Techkon SpectroDens v4
NIX Spectro 2  soon

 

Some of the differences between instruments are due to measurement devices measuring different spectral ranges, but the most critical difference is due to differences in their Light source (illuminants) and monochromator.

 Instrument Light
Source
Monochromator
Detector
Comments
i1 Pro 3  Full-Spectrum LED  diffraction grating
128-pixel diode  array
 
i1 Pro 2  Gas-filled tungsten (illuminant A)
and UV LED

diffraction grating
128-pixel diode  array

 
eXact Gas-filled tungsten (illuminant A)
and UV LED
DRS spectral engine
Two different calibrations for simultaneous M-condition measurement. M1 (part 1), M1 (part 2)  
Techkon SpectroDens v4  Full-Spectrum LED diffraction grating  
NIX Spectro 2  Full-Spectrum LED diffraction grating  

 Additionally. eXact in version 1 has two modes (M1 Part 2 Export on or off)

 

 

 

 

 

 

 

 

 