Support

FAQ

    June 5, 2024

    Calibrate ColorChecker Studio (Xrite i1Studio)

    Calibrate ColorChecker Studio replaces the X-Rite i1Studio (EOSTUDIO)

    • M2 measurement Conditions (UV cut)
    • Spectral reporting: 10 nm, between 380 nm, and 730 nm
    • Aperture 6 mm
    • Max Luminance measurements 1000 nits
    • Short Term Repeatability Emissive x, y: ±0.002 typical (5000K @ 80 cd/m²)
    • Short Term Repeatability Reflective 0.1ΔE*2000 (D50, 2°) typical, mean value of 10 measurements on white
    • Inter-instrument agreement: 0.6ΔE*2000 average, max. 1.5ΔE*2000 max. (deviation from X-Rite i1Studio manufacturing standard at a temperature of 23°C for single measurement mode on 12 BCRA tiles D50, 2°)
    • Light Source White LED

    Modes:

    • Spot Color Measurements (Color Inspector)
    • Scan Control Strips (Print Inspector)
    • Lighting Measurements (Lighting Inspector)
    • Display Calibration and profiling (Display Inspector)

    Important Instrument Limitations:

    • Patch size 20 x 29 mm + 2 mm separation bar
    • max 15 patches in a single scanning row (328 mm)
    • No OBAs Index – instrument can't quantify the presence of optical brighteners
    • No Fluorescence Index/Fluorescent Effect evaluation

    Please note:

    This instrument is dedicated to Enthusiasts and Professional Photographers who print, Graphic/Packaging/Industrial/Interior Designers, and Stylists. It supported the M2 measurement condition only – which might be a limitation for some professional usage areas.  

     